Matrix
hydrophilic polymer bead
hydrophilic polymer bead
Pore size(nm)
100
100
Particle size(μm)
5
5
Charged group
-CH2N+(CH3)3
-CH2CH2CH2SO3
Counter ion
Cl-
Na+
Ion-exchange capacity
0.075-0.100 meq/ml-resin 0.070-0.095 meq/ml-resin
Dynamic binding capacity
110-150 mg-BSA/ml-resin 70-100 mg-human IgG/ml-resin
Available pH range
2.0 - 12.0 2.0 - 12.0
 
 
High binding capacity and high recovery
The porous type of YMC-BioPro shows great absorption capacity and excellent recovery, and is especially suitable for semipreparative separations of proteins and antibodies.
 
Column Dimension Part Number
YMC-BioPro SP 50x4.6 mm ID
SPA0S050546WP
YMC-BioPro QA 50x4.6 mm ID
QAA0S050546WP
 
 

Comparison of dynamic binding capacity (DBC) for BSA
Compared with conventional porous polymer anion-exchange columns, YMC-BioPro QA gives
higher DBC and recovery rates. This indicates that YMC-BioPro has a much lower nonspecific
adsorption compared to conventional columns.

 
  Dynamic binding capacity
(mg/ml-gel, 10% breakthrough)
Eluted amount
(mg/ml-gel)
Recovery* (%)
YMC-BioPro QA
126
120
95
Mono Q
100
35
35
BioAssist Q
73
58
79
*Recovery: (Eluted amount/Dynamic binding capacity) x 100
 
YMC Biopro IEX Columns Chromatogram
 

Before the determination, equilibrate the column with equilibration buffer.


Step 1: Load
A protein solution of known concentration is continuously loaded at the desired fl ow rate and the absorbance of the eluate is monitored until full saturation is achieved (100% UV absorbance of the pure sample solutions).
Step 2: Wash
Step 3: Elute
- Determination of DBC
The volume of protein solution that has been applied up to a specific breakthrough point (usually 5 or 10%) is a measure of the DBC of the medium.

High Mass Recovery
The high purity silica support and complete bonding coverage of Protein-RP results in very high sample recoveries. This is an important consideration when choosing a column for biochemical analyses. The table below shows recoveries of Proteins and Peptides. YMC-Pack Protein-RP provides higher recoveries than competitive C4 columns.
 
 
Superior resolution

YMC Biopro IEX Columns
Eluent: A) 20 mM KH2PO4-K2HPO4 (pH 6.8)
B) 20 mM KH2PO4-K2HPO4 (pH 6.8)
containing 0.5M NaCl
0-100% B (0-60 min., linear)
1. Cytochrome c (0.5 mg/ml)
2. Ribonuclease A (0.5 mg/ml)
3. Lysozyme (0.5 mg/ml)
Flow rate: 0.5 ml/min (4.6 mm i.d. column)
0.59 ml/min (5.0 mm i.d. column)
 
Temperature: 25 °C  
Detection: UV at 220 nm  
Injection: 20 μl (4.6 mm i.d. column)
23.6 μl (5.0 mm i.d. column)
 
*By courtesy of YMC Co., Ltd.
 
  

YMC Biopro IEX Columns

 


 
 
©Crawford Scientific 2010