YMC-Pack Pro C18

YMC-Pack Pro C18 is based on an ultra pure silica support which is used for the whole ProFamily. Due to proprietry andcapping this column is perfectly suitable for the separation of acidic and basic molecules. The inertness of the silica makes it an excellent choice for the analysis of drugs and their metabolites as well as compounds that are susceptible to polar interactions with residual silanol groups and metals.

Extremely basic substances are selected to prove the performance of YMC-Pack Pro C18 with regards to their separation and peak performance which cannot be acheived with classical materials.

Separation of extremely basic compounds
YMC-Pack Pro C18 - Fig 3
Tricyclic antidepressants
Column dimensions: 4.6 x 150 mm, 5 µm
Eluent: 20 mM potassium phosphate (pH 6.9):methanol (25:75)
Flow rate: 1.0 mL/min
Temperature: 37°C
Detection: UV at 254 nm
  • Specifically designed for research and development in the pharmaceutical and biotechnology industries
  • Extremely narrow specifications
  • High lot-to-lot reproducibility
  • High column-to-column reproducibility
  • Ideal for basic, acidic and polar compounds
Specifications
  YMC-Pack Pro C18
Particle size (µm) 2, 3, 5
Pore size (nm) 12
Surface area (m2/g) 330
Carbon content (%) 16
Recommended pH range 2.0-8.0
Reproducibility

High lot-to-lot reproducibility is the essential basis for high performance HPLC columns.

High column performance can only be achieved if a column is well packed to give continuously high numbers of theoretical plates and therefore good asymmetry factors. The figure below clearly demonstrates YMC's high level of reproducible quality from column to column.

Column-to-column comparison for consecutive packed columns
YMC-Pack Pro C18 - Fig 2
Column dimensions: 4.6 x 150 mm, 5 µm
Mobile phase: 60:40 acetonitrile:water
Flow rate: 1.0 mL/min
Detection: UV at 270 nm, 0.32 AUFS
Temperature: 25 ºC
Plate count and asymmetry were calculated for butyl benzoate.
Column care
YMC Pack Pro C18 is stable towards hydrolysis between pH 2-8. Remove acid and buffer salts before storage. Store the column in 70:30 methanol:water.
For detailed information please refer to the “Column Care and Use Instructions” which are shipped with each analytical column.
Ordering Information
3 µm
12 nm
Column Length (mm)
Column I.D. (mm) 30(WT)/
33(QT)
50 100 150 250

Guards
(10 mm) 5/pk

2.1 AS12S03-H3Q1QT AS12S03-05Q1QT AS12S03-10Q1QT AS12S03-15Q1QT AS12S03-25Q1QT AS12S03-01Q1GC
3.0 AS12S03-H303QT AS12S03-0503QT AS12S03-1003QT AS12S03-1503QT AS12S03-2503QT AS12S03-0103GC
4.0 AS12S03-H304QT AS12S03-0504QT AS12S03-1004QT AS12S03-1504QT AS12S03-2504QT AS12S03-0104GC
4.6 AS12S03-0346WT AS12S03-0546WT AS12S03-1046WT AS12S03-1546WT AS12S03-2546WT AS12S03-0104GC

5 µm
12 nm
Column Length (mm)
Column I.D. (mm) 30(WT)/
33(QT)
50 100 150 250

Guards
(10 mm) 5/pk

2.1 AS12S05-H3Q1QT AS12S05-05Q1QT AS12S05-10Q1QT AS12S05-15Q1QT AS12S05-25Q1QT AS12S05-01Q1GC
3.0 AS12S05-H303QT AS12S05-0503QT AS12S05-1003QT AS12S53-1503QT AS12S05-2503QT AS12S05-0103GC
4.0 AS12S05-H304QT AS12S05-0504QT AS12S05-1004QT AS12S05-1504QT AS12S05-2504QT AS12S05-0104GC
4.6 AS12S05-0346WT AS12S05-0546WT AS12S05-1046WT AS12S05-1546WT AS12S05-2546WT AS12S05-0104GC
** Guard cartridge holder required is part number XPGCH-Q1

 
 

 

 

 

   
 
 
YMC ProFamily
 

 
 