We seem to have entered something of a ‘mixed-mode’ world in reversed phase HPLC. Many stationary phases are now available which incorporate permanent or induced charges within, or alongside, hydrophobic alkyl silica type bonded ligands. 

Whilst the retention characteristic of these phases adds a dimension of complexity when planning, developing or troubleshooting separations, the extra capabilities that these phases offer within the reversed phase separation space can be highly beneficial. 

There are important distinctions to be made when considering phases in which the charge can be manipulated using pH adjustment of the eluent and those which are permanently charged, regardless of eluent pH.  Stationary phases where the degree of surface charge can be manipulated using eluent pH, offer greater ‘tunability’ of selectivity, but are more complex in terms of the requirements for accurate pH control, column (re)equilibration and method development. 

Phases which are permanently charged, can feature the charged moiety within the alkyl chain (either closer to or further away from the surface), or by direct application to the silica surface. 
 In recent years, a category of stationary phases colloquially known as ‘Charged Surface’ phases have been introduced to offer several chromatographic advantages, these phases having a ‘balanced’ charge which is sufficient to overcome several important chromatographic issues, especially when using MS detection.  These phases are designed to undergo lower energy interactions with ionogenic analytes when compared to the so-called ‘mixed-mode’ phases in which higher energy ion-exchange interactions are used to significantly alter analyte retention and selectivity. 

Figure 1: Depiction of approaches to achieving a Charged Surface type C18 stationary phase 

Figure 1 A and B show a bonded C18 alkyl ligand alongside an uncharged ‘lone’ silanol group, which remains post ligand bonding and can undergo dipole interactions with polar analytes.  When the pH of the eluent is raised above around pH 6-7, these silanols groups can ionise and the resulting anionic surface can undergo significant secondary interaction with charged basic analytes, often leading to significant peak shape deformation.  To combat this, so called ‘Charge Surface’ columns are modified with a positive charge which helps to shield ionogenic basic analytes from the silica surface, improving both peak shape and loadability when using low ionic strength eluents.  The mechanism of achieving the surface charge is proprietary for all manufacturers, but potential methods include the use of a bonded, ionised, end capping reagent as shown in Figure 3D and the direct modification of the silica surface by crosslinking with a reagent which includes an ionogenic moiety as shown in Figure 3A. 

These ‘charged surface’ phases offer several advantages to the chromatographer including;

  1. Alternative selectivity
  2. Improved peak shape and loading for basic analytes in low ionic strength mobile phases
  3. Reduced re-equilibration times when switching from high to low pH columns 

Of the listed advantages, perhaps the most important are peak shape and loadability improvements, driven mainly by modern approaches to eluent design when using MS detection.   

To maximise sensitivity with MS detection, chromatographers are choosing to reduce the ionic strength of eluent systems or simply to alter pH using weaker acids or bases and omit the use of a salt counter ion.  Whilst this approach may leave the analysis more susceptible to changes in retention time and selectivity, this is often outweighed by the improvements in detector sensitivity through the lowering of charge competition within the sprayed droplets in electrospray atmospheric pressure ionisation.  



Poor peak shape, especially for charged basic analytes is a well-documented phenomenon and is caused by several factors, especially in low ionic strength eluent systems [1-3].  The interaction of the cationic analyte with surface anionic sites, predominantly charged surface silanol groups, can cause secondary interactions which result in peak tailing.  Whilst the use of high purity silica and advanced silica bonding and deactivation (end-capping) regimes have reduced these phenomena, it can remain problematic.  The loadability of stationary phases for charged analytes in low ionic strength eluents can be orders of magnitude less than neutral analytes and peak tailing can occur even with very low loadings of charged analytes, depending upon the ionic strength of the eluent, the surface area of the packing material, extent of analyte ionisation and the use and concentration of a counter ion (ion-par), such as trifluroracetic acid etc. 

The use of an intentionally introduced, low level, positive charge on the silica surface can overcome secondary interactions with surface silanol groups, essentially by providing a repulsion effect towards cationic analytes, i.e. a basic analyte at eluent pH significantly below the pKa value.  The mechanism for increased loadability of the charged surface phases is not well understood but can be demonstrated empirically through the application of varying degrees of surface charge.  With too little surface charge, peaks for charged basic analytes at low eluent pH tend to suffer from bi-Langmurial, or fronting, peak behaviour and with too great a surface charge they suffer from anti-Langmurial, or tailing behaviour, suggesting an optimum surface charge is required to produce Gaussian, neither fronting nor tailing, peak shapes.  When analysing charged basic analytes at low concentrations, peak shape is important to take advantage of other improvements in column efficiency such as reduced dimension particles. 

Whilst the extent and strength of ion-exchange behaviour in these charged surface phases is intended to be lower than that used in more traditional mixed-mode type phases, there can, nevertheless, be an impact on the selectivity of the stationary phase.  This depends upon the degree of surface charge, the nature of the bonded phase ligand (and therefore the accessibility of the charged silica surface), the extent of ionisation of the analyte molecules and their charge state (cationic or anionic).  It is important to highlight that whilst a cationic surface charge is convenient in the repulsion of basic analytes and therefore their improved peak shape, these charges may also act to increase retention for any charged acidic analytes present in the sample.  

To further highlight this point, a comparison was made between various columns of nominal L1 classification (C18) to investigate the relative retention of acidic, neutral and basic species on traditional C18 and Charged Surface C18 phases.  We used a recently updated version of the Hydrophobic Subtraction Model (HSM2) [3] to compare relative retention of acidic and basic analyte test probes using calculated (simulated) chromatograms.  The model uses ethyl benzene retention time as a comparator for each test probe and uses these relative retention times to describe various interactions between analyte and stationary phase including  hydrophobic retention, hydrogen bonding, cation exchange, accessible volume and dipole interaction according to Equation 1; 

[1] 
h (H)- solute / column hydrophobicity  
b (A) – solute / column hydrogen bond basicity 
a0 (B) – solute / column hydrogen bond acidity  
k (C) – solute / column cation exchange ability (protonated bases at pH 2.8)   
v (V) – solute / column accessible volume (analyte hydrodynamic volume) 
d (D) – solute / column dipole parameter 
 

The columns chosen for the comparison were from the same manufacturer with the same stationary phase particle size, the same underlying silica and the same bonded phase ligand (C18) thus ensuring that any differences in analyte retention were predominantly due to the applied surface charge. 

Column parameters (H, A, B etc.) for each of the analyte / stationary phase interaction descriptors can be made using empirical measurements of a variety of test probes with carefully chosen physico-chemical properties.  The values of these column parameters for the test columns chosen are shown in Figure 2. 

Figure 2: Relative magnitude of each analyte / stationary phase interaction terms within the revised Hydrophobic Subtraction Model for a C18 bonded column and a charged surface variant of the same column 

It is interesting to note, amongst other differences between the phases, that the Cation Exchange capacity (C Term) of the charged surface stationary phase is considerably lower, in keeping with the theory that the positively charged surface shields basic analytes from interaction with any anionic residual silanol groups on the silica surface.  Further, the hydrogen bond acidity (A Term) of the phase describes the ability of non-ionised silanol residues to interact with non-ionised basic analytes.  Again, the lower value of this parameter is consistent of the shielding (repulsive) effect of the positive surface charge preventing these interactions. 

In the absence of the original screening retention times for the columns within the database, it is possible to use the HSM equation values for analytes and stationary phases to reconstruct a simulated chromatogram of the 16 test probes of the Hydrophobic Subtraction Model.  These simulated chromatograms are shown in Figure 3.  

Figure 3:  Simulated chromatograms for C18 (top chromatogram) and C18 Charged Surface (lower chromatogram) Columns derived from the revised Hydrophobic Subtraction Model.  A:  Full Chromatogram for all 16 test probes, B: Charged Basic Analytes, C: Charged Acidic Analytes 
 
The relative retention factors for the acidic and basic analytes highlighted in Figure 3 are shown in Table 1, alongside the relative magnitude in the retention factor change. 

Table 1: Retention Factors for analytes in simulated chromatograms from Figure 3 and % change for each analyte between C18 and C18 Charged Surface phases 

It is interesting to note that the relative changes in retention behaviour fit well with the theoretical behaviour of charged surface columns, i.e. a decrease in the retention of basic species due to the slight repulsive effects of the cationic surface charge and an increase in the retention of acidic analytes due to an ion exchange interaction with the cationic surface charges. 

It should be noted, that in deriving the HSM2 parameters, the initial test probe retention time measurements were undertaken at approximately neutral pH with a 60mM phosphate buffer.  Whilst the acidic and basic analytes shown in Figure 3 will have been fully ionised at the experimental pH, the effects of the higher buffer concentration on analyte / stationary phase interactions on the charged surface cannot be predicted.  This being said, we have compared a number of C18/ Charge Surface C18 columns pairs using this approach and all have shown the same behaviour, with a slight decrease in retention of charged basic compounds and an increase in retention of charged acidic compounds. 

It is clear from Figure 3 (Top) that the simulated test probe chromatograms for the C18 and the charged surface equivalent show significant selectivity differences, and this should be borne in mind by chromatographers wishing to take advantage of the improved peak shape of charged basic analytes in low ionic strength mobile phases. 

In terms of improved peak shape of basic analytes, much is demonstrated in column manufacturer literature to show that this claimed advantage is bourn out in practice and is therefore an advantage for chromatographers wishing to analyse charged basic analytes in low ionic strength mobile phases and to attain the maximum sensitivity. 

References

  1. McDonald, P.D. and Bidlingmeyer, B. Chapter 1 in: Bidlingmeyer, B, Ed. Preparative Liquid Chromatography; Elsevier, New York, 1986, pp. 37-43 
  2. Snyder, L Principles of Adsorption Chromatography, Elsevier, Marcel Dekker, New York, 1967, pp. 77-97 
  3. Guichon, G., Shirazi, S.G., Katti, A., Fundamentals of Preparative and Non-Linear Chromatography; Academic Press; Boston, Ma, 1994. 
  4. Stoll, D.R., ∗, Dahlseid, T.A., Rutan, S.C., Taylor T. , Serret, J. M., Improvements in the predictive accuracy of the hydrophobic subtraction model of reversed-phase selectivity, Journal of Chromatography A 1636 (2021) 461682 

Tony Taylor, Chief Scientific Officer, Element Life Sciences (EMEAA)

Tony has worked as a chromatographer in the Pharmaceutical, Polymer, Contract Analysis and Consulting Industries for more than 35 years.  His experience includes; HPLC and GC method development, development and troubleshooting of LC-MS and GC-MS methods, HPLC stationary phase characterisation, targeted and untargeted trace analysis, GC-MS spectral interpretation, solid phase extraction and development of sample preparation methods.  Tony is a founder of  CHROMacademy and has delivered training in chromatographic analysis to thousands of students globally.